Calibration settings for BioSonics data

This page describes the settings, necessary to an Echoview calculation, on the Calibration page of the Variable Properties dialog box for acoustic variables derived from Biosonics data files.

We recommend you assess the suitability of the settings available on the Calibration pages for each acoustic variable. Some settings are used to calculate datapoint values or ranges, others may be used during analysis or display of the data. The table below shows which settings are used for calculating datapoint values and ranges, which are used for other calculations and which settings may not be used by the variable.

Quantitative calculation/analysis requires calibrated data which is achieved using correct calibration setting values (rather than default values). An ECS file enables you to modify calibration settings for variables in a fileset.

Note: The (red) default values on the Calibration pages are for illustration purposes only and have not been read from your data file. Read About calibration settings for further information.

Table legend:

P

Setting is used in calculating datapoint values or ranges as indicated by the column heading.

 

Setting is not used in calculating datapoint values (when green background) or ranges (when blue background) but may be used for other purposes, see Additional uses of Calibration settings

 

Setting is not available for this variable

Setting

Datapoint values

Datapoint range

Sv

TS

Angular positions

All acoustic variables

Calibration page of Variable Properties dialog box

AbsorptionCoefficient

P

P

 

 

CalibrationOffset

P

P

 

 

Dt4PowerSetting

P

P

   

Frequency

 

 

 

 

MajorAxis3dBBeamAngle

P

     

MajorAxisAngleOffset+

 

 

P

 

MinorAxis3dBBeamAngle

P

     

MinorAxisAngleOffset+

 

 

P

 

SoundSpeed

P

 

P

P

TransmittedPulseLength

P

 

 

 

TvgRangeCorrection

P

Not used for the TS variable.

   

TwoWayBeamAngle1

 

 

 

 

WedgeVolumeBeamAngle

 

     

From data file

InitialBanking

 P

 P

 

P

SamplePeriod

 P

 P

 

P

SamplesPerPing

P

P

 

P

Notes:

Other required calibration settings are read from the BioSonics data file (EEPROM structure).

See also

About calibration settings
Entering calibration settings

BioSonics files

Raw variables derived from Biosonics data