Calibration page
The Calibration page of the Variable Properties dialog box or the Variable Properties - Multiple Variables Selected is displayed for acoustic variables and is used to calibrate data for the variable.
We recommend you assess the suitability of the settings available on the Calibration pages for each acoustic variable. Some settings are used to calculate data point values or ranges, others may be used during analysis or display of the data. The calibration setting can change with Echoview calculations.
Quantitative calculation/analysis requires calibrated data which is achieved using correct calibration setting values (rather than default values). An ECS file enables you to modify calibration settings for variables in a fileset. See also ECS example, calibration values used by Echoview.
The calibration settings can affect:
- The calculation of the data values for a variable; and/or
- Sample range values (check the validity of existing regions and lines after changes); and/or
- Processing outputs
Check the calibration page of each variable you are working with.
- A calibration page must be completed for each variable you are working with as the properties of all variables are independent.
- It is advised that any transducer, referred to on the calibration page, be configured (see Transducer Properties dialog box).
See About calibrated data and About calibration settings for further explanation.
The Calibration page sections for both raw and virtual variables include:
Additional information:
Setting |
Description |
||||||||
Select a Transducer from the Transducer list. A Transducer object in the Dataflow window allows you to configure a transducer. The Transducer Properties allow you to record general information, and enter settings for transducer location and orientation. |
|||||||||
Available only for virtual variables. Select Override transducer inheritance when you want to use a transducer other than the transducer associated with the first operand. This setting may change when you drag and drop from a transducer node on the Dataflow window. Note: EV files for Echoview 6.0 and earlier are updated with respect to the transducer of the first operand. |
|||||||||
Edit the calibration file used by this fileset |
This hyperlink is only active when an ECS file has been added to the fileset. Note: An ECS file may be used by multiple EV files or multiple variables. For more information refer to Echoview Calibration Supplement files. |
||||||||
SourceCal |
Refer to the SourceCal entry. |
||||||||
LocalCal |
Refer to the LocalCal entry. |
||||||||
Settings |
Displays the calibration settings that Echoview may use. Echoview handles an extensive range of calibration settings, a complete list (including default values, units and ranges) can be found under Help menu > Calibration Information. Only a subset of these are relevant to your data. Echoview knows a setting is relevant when it reads it from a data or ECS file, or uses it in a calculation. These are the ones that were read from the variable's own calibration source in the data file, or explicitly associated with the fileset or calibration source in the ECS file. The Variable Properties Calibration page only displays the calibration settings that are relevant to the variable's data. The Used by pings section displays the ping range generated with the displayed calibration settings. The Value column lists the values used in the calculation. The Calibration file column lists the values read from the ECS file in the fileset. The Data file column lists the values read from the data files in the fileset.
For a technical description of the various calibration settings please consult: Notes:
|
The Sonar calculator (Help menu > Sonar Calculator) can calculate Sound speed and Absorption coefficient in sea water or fresh water using well known algorithms.
Raw variables
Available settings by format
For details on the calibration settings for your equipment please refer to the relevant section below.
-
Calibration settings for text data
Calibration settings for ASL Environmental Sciences data
Calibration settings for BioSonics data
Calibration settings for ARIS data
Calibration settings for Blueprint Subsea Oculus data
Calibration settings for BlueView data
Calibration settings for DIDSON data
Calibration settings for EchoListener data
Calibration settings for Echoview Data File Format data
Calibration settings for Furuno FQ80 data
Calibration settings for Furuno FCV30 data
Calibration settings for Furuno FCV38 data
Calibration settings for Furuno ETR-30N data
Calibration settings for Furuno FSV-25 (netCDF) data
Calibration settings for Furuno FSV-30 (Research version) data
Calibration settings for HAC data
Calibration settings for HTI data
Calibration settings for Kongsberg EM 3002 and EM 3002D data (*.all)
Calibration settings for Kongsberg EM 710 data (*.all)
Calibration settings for Kongsberg EM systems (*KMall)
Calibration settings for Kongsberg Flexview .mmb data converted to the .imb file format
Calibration settings for Kongsberg Mesotech MS 1000 data
Calibration settings for Kongsberg M3 data
Calibration settings for Kongsberg EA640
Calibration settings for Navico Group Lowrance data
Calibration settings for Norbit data
Calibration settings for Nortek data
Calibration settings for PAS data
Calibration settings for R2Sonic data
Calibration settings for RD Instruments workhorse data
Calibration settings for RESON 6012 data
Calibration settings for RESON SeaBat 7000 series data
Calibration settings for RESON Seabat T20-P data
Calibration settings for RESON Seabat T50-R data
Calibration settings for RESON SeaBat 8000 series data
Calibration settings for Teledyne Odom data
Calibration settings for Simrad Omnisonar netCDF data
Calibration settings for Simrad EK15 data
Calibration settings for Simrad EK80 raw data
Calibration settings for Simrad EK80 EC150-3C data
Calibration settings for Simrad EK500 telegrams logged with Ex500
Calibration settings for Simrad EK500 telegrams logged with Ex60
Calibration settings for Simrad Ex60 raw data
Calibration settings for Simrad Ex70 raw data
Calibration settings for Simrad SM2000 data
Calibration settings for Simrad SM20 data
Calibration settings for Simrad ME70 raw data
Calibration settings for Simrad MS70 raw data
Calibration settings for Simrad SH80 data
Calibration settings for Simrad SH90 data
Calibration settings for Simrad SC90 data
Calibration settings for Simrad SP70 data
Calibration settings for Simard omnisonar netCDF data
Calibration settings for Simrad SX90 data
Calibration settings for Sonic data
Calibration settings for WASSP data
See also
Entering Calibration Settings
Variable Properties dialog box
Virtual variables
Virtual variables, like raw variables, must be correctly calibrated to ensure meaningful analyses. Display the Dataflow window to create and use virtual variables. Calibration settings are displayed on the virtual variable's Calibration page of the Variable Properties dialog box.
Virtual variables display SourceCal settings and any applied LocalCal settings. Changes to the SourceCal (up the virtual chain) flow onto the virtual variable. Virtual variable LocalCal behavior is discussed under Inheritance of Properties. In addition, virtual variables may use particular calibration settings for display, analysis or other purposes.
Settings that may be used by a virtual variable
- AbsorptionCoefficient
- SoundSpeed
- PulseDuration
- TwoWayBeamAngle
- CalibrationOffset1
- TvgRangeCorrection
- TvgRangeCorrectionOffset
- MajorAxis3dBBeamAngle, MinorAxis3dBBeamAngle
These settings may also be used for calculations in analyses or for other purposes. For more information on this, see Additional uses of calibration settings.
Notes:
- The 3dB beam angle setting is generally not used for virtual variable data point calculations. However, for Single target and Target length virtual variables this setting may be important.
- TvgRangeCorrection may be specified or applied by default for some data formats.
Settings that may be used in specific analyses of a virtual variable
These settings can affect specific analyses i.e. operations such as calculating schools analysis variables.
For more information on this, see Additional uses of calibration settings.
Notes:
- The 3dB beam angle setting does not appear in this list. However, for Sv and Single target virtual variables this setting may be important.
- The Thickness source on the Data page of the Variable Properties dialog box does not appear in this list, however, the setting may be important to single target and fish track work. For more information, refer to About Single targets: Configuring the display.
- Specific additional notes describing how the calibration settings are used by the virtual variable operators are available on the following pages:
- When you Copy-Paste TVG range correction calibration settings (to an ECS file), be aware that if the value is variable-specific it may not apply to other variables that use the same Calibration source. See for TVG range correction settings for more details.
See also
About calibration settings
About virtual variables
Creating and deleting virtual variables
Entering calibration settings
Inheritance of properties
About calibrated data
Calibrating an echosounder
Beam angle parameters and related calculations in Echoview